Question of the Day:

I want to buy a new mammography unit, but it has a tungsten target. Don't I need the characteristic x-rays from Molybdenum to have the optimal energy range for breast imaging?

Tip of the day:

Dose alerts are set for equipment as a complete unit. This means it may help prevent overdosing a patient, but it also means that the alert may kick in during a high-dose procedure like CT-fluoroscopy and interrupt imaging. Always make sure someone in the room has the password to override when performing high dose procedures on dose alert enabled equipment.